Pioneering technologies in fiber optic fusion splicing

Designed for the highest demands - advanced systems for fiber optic preparation, testing and analysing.

CleaveMeter 2



Fiber Optic Proof Tester

Short facts

  • Sharp fringe patterns
  • Flat and angled cleave measurements with in-picture results
  • Operator skill independent for fast operation in production environments
  • Documents images from critical end-faces in production, research and development
  • Software compensation for fiber holder error
  • Three-point fiber diameter measurement
  • Accepts fibers with cladding diameters from 125 µm to 1200 µm
  • Accepts fiber holders of major splicer manufacturers

Description

The CleaveMeter 2™ is a non-contact interferometer designed for inspecting the end-faces of cleaved, lensed or polished optical fibers with cladding diameters of 125 µm to 1200 µm. It gives immediate information on important end-face properties such as flatness, perpendicularity, hackles and dust. Sampling tests as well as continuous process documentation can be carried out both easily and quickly, making this an ideal instrument for cleaver inspection and optimization.

By means of the host application the operator can view the live camera image, adjust camera settings, save and load images to and from file, and log all information in a specific text-based log file. Software-compensation for fiber holder error allows for high-precision measurements with in-picture presentation of the cleave angle. This enables operator skill independence and easy process verification in production environments. Other useful software features include user-defined markers at points of interest, three-point fiber diameter measurement and pseudo-color mode for better contrast.

The optical system is based on a high-end camera with true megapixel resolution and very high sensitivity, yielding excellent image quality at high frame rates and high magnification. Switching between low and high magnification is software-controlled. High-precision optics guarantees sharp and clear pictures with very little aberration.

Adaptor plates are available for both perpendicular and angled cleave measurements.The mechanical design is compatible with all NYFORS automatic fiber cleavers and accepts the fiber holders used with those machines as well as those of major splicer manufacturers. Customized adaptor plates are available upon request.

The CleaveMeter 2™ comes in a small ergonomic, bench-top design and connects to the USB port of a PC running the host application.


Technical Data

Fiber cladding 125-1200µm
Fiber coating 250-1500µm
Camera resolution 1280x1024 pixels
Image scale 1.25 µm per pixel
Image file format 8-bit JPEG, PNG, TIFF, BMP
PC connection USB 2.0 port
Power supply Through USB port
Dimensions 97 mm (W) x 179 mm (D) x 142 mm (H)
Weight 1.6 kg

NYFORS part number: 30000020

Included in delivery: CleaveMeter 2™ unit, PC Software, USB interface cable, Manual and Tools.

Cleave Angle Accuracy

Absolute accuracy <0.03 degrees*
Relative accuracy 20 % of measured cleave angle (125-199 µm)
10 % of measured cleave angle (200-529 µm)
5 % of measured cleave angle (530-1200 µm)

*Fiber holder error not included. Adaptor plates are delivered with individual certificates for precise software compensation of fiber holder error in cleave angle measurements. For more information about system accuracy, please contact us at This e-mail address is being protected from spambots. You need JavaScript enabled to view it .

Selection Guide

Fiber-specific adaptor plates are required in order to properly align different fiber sizes to the center of the CleaveMeter™ optical system field of view. They are not included in delivery and should be ordered separately.

Adaptor plates are available for use with NYFORS automatic fiber cleavers and fiber holders of major splicer manufacturers such as Ericsson, Fujikura/AFL and Furukawa/FITEL. Below you find a selection of the most common types and dimensions. NYFORS generic adaptor plates are compatible with NYFORS LD fiber clamps and Ericsson FSU-clamps.

Please select adaptor plate to match fiber cladding diameter and angle adaptor plate (optional) to match fiber tilt angle.

For more information about available adaptor plates and custom sizes, please contact us at This e-mail address is being protected from spambots. You need JavaScript enabled to view it .


Adaptor Plate

Cladding diameter (µm) Article Description Article number
Customer specified Adaptor plate, NYFORS, Generic 30020010
Customer specified Adaptor plate, FJK, Generic 30010000
115-210 Adaptor plate, FJK, 115-210 µm 30010005
200-529 Adaptor plate, FJK, 200-529 µm 30010010
510-800 Adaptor plate, FJK, 510-800 µm 30010015
800-1200 Adaptor plate, FJK, 800-1200 µm 30010020

Angle Adaptor Plate (optional)

Fiber tilt angle
Customer specified Angle adaptor plate, Generic 30040000
8 degrees Angle adaptor plate, 8 degrees 30040080
15 degrees Angle adaptor plate, 15 degrees 30040150

CleaveMeter 2 Product Sheet